1 Non-Contrast Scan              10+ Actionable Reports

Opportunistic AI-enabled automated bone mineral density measurements

  • May 19, 2023
  • 2 min read

Updated: Nov 11, 2024

in lung cancer screening and coronary calcium scoring CT scans are equivalent


Naghavi M, De Oliveira I, Mao SS, Jaberzadeh A, Montoya J, Zhang C, Atlas K, Manubolu V, Montes M, Li D, Atlas T, Reeves A, Henschke C, Yankelevitz D, Budoff M. Opportunistic AI-enabled automated bone mineral density measurements in lung cancer screening and coronary calcium scoring CT scans are equivalent. Eur J Radiol Open. 2023 May 13;10:100492. doi: 10.1016/j.ejro.2023.100492. PMID: 37214544; PMCID: PMC10196960.

European Journal of Radiology. 05/13/2023


Abstract

Rationale and objectives: We previously reported a novel manual method for measuring bone mineral density (BMD) in coronary artery calcium (CAC) scans and validated our method against Dual X-Ray Absorptiometry (DEXA). Furthermore, we have developed and validated an artificial intelligence (AI) based automated BMD (AutoBMD) measurement as an opportunistic add-on to CAC scans that recently received FDA approval. In this report, we present evidence of equivalency between AutoBMD measurements in cardiac vs lung CT scans.

Materials and methods: AI models were trained using 132 cases with 7649 (3 mm) slices for CAC, and 37 cases with 21918 (0.5 mm) slices for lung scans. To validate AutoBMD against manual measurements, we used 6776 cases of BMD measured manually on CAC scans in the Multi-Ethnic Study of Atherosclerosis (MESA). We then used 165 additional cases from Harbor UCLA Lundquist Institute to compare AutoBMD in patients who underwent both cardiac and lung scans on the same day.


Results: Mean±SD for age was 69 ± 9.4 years with 52.4% male. AutoBMD in lung and cardiac scans, and manual BMD in cardiac scans were 153.7 ± 43.9, 155.1 ± 44.4, and 163.6 ± 45.3 g/cm3, respectively (p = 0.09). Bland-Altman agreement analysis between AutoBMD lung and cardiac scans resulted in 1.37 g/cm3 mean differences. Pearson correlation coefficient between lung and cardiac AutoBMD was R2 = 0.95 (p < 0.0001).

Conclusion: Opportunistic BMD measurement using AutoBMD in CAC and lung cancer screening scans is promising and yields similar results. No extra radiation plus the high prevalence of asymptomatic osteoporosis makes AutoBMD an ideal screening tool for osteopenia and osteoporosis in CT scans done for other reasons.




Comments


Ready to Unlock the Value in Your CT Scans?

Join leading healthcare providers using AI-CVD to save lives and increase revenue, and lead the future of preventive cardiology.

  • Twitter
  • LinkedIn
  • Facebook
  • Instagram
  • Youtube

2450 Holcombe Blvd
TMC Innovations
Houston, TX 77021

aicvd patent_edited_edited.jpg
ISO LOGO_edited.png

©2026 HeartLung Corporation. All Rights Reserved. US Patent Nos US9119590*, US10695022, US11610686, US18167691, US20240115224, and Patents Pending. AI-CVD, AI-CAC, AutoBMD, AutoCAC, AutoChamber and other trademarks shown on this website are protected under intellectual property rights of HeartLung Corporation in the United States.